The HK-18854-L2 is a benchtop automatic oil particle counter for contamination analysis of hydraulic oil, lubricating oil, fire-resistant oil, insulating oil, turbine oil and other petroleum-based fluids. It uses the light-blockage method and is designed to meet ISO 11171 / GB/T 18854 requirements.
Key features
- 0.8–600 µm particle-size range
- 8–64 detection channels, 64 channels standard
- Sampling volume 0.2–6000 mL
- Sampling accuracy better than ±0.2%
- High-viscosity sample capability up to 650 cSt
- Positive and negative pressure chamber for degassing and high-viscosity samples
- Built-in standards including ISO 4406, NAS 1638, SAE and GJB
- 7-inch colour touch screen
- RS232, RS485 and USB connectivity
- Optional temperature, water activity and moisture measurement modules
Technical data
- Repeatability: RSD <2%
- Sampling speed: 5–80 mL/min
- Test sample temperature: 0–80 °C
- Operating temperature: -20–60 °C
- Power: AC 110–245 V, 50/60 Hz
- Main unit: 340 × 410 × 650 mm
- Main unit weight: ≤23 kg
Datasheet: Download Datasheet (PDF)