DESCRIPTION
PCE-VC20 is a vibration meter calibrator used to calibrate vibration sensors, analyzers and testers. Suitable for calibrating equipment weighing up to 600 g (1.32 lb), this vibration calibrator generates mechanical vibrations with a quartz-stabilized vibration frequency of 159.15 Hz (radian frequency 1000/s).
A control circuit and reference accelerometer inside the excitation or shaker head work in tandem to keep the vibration level constant. A display indicates the error percentage (%). The error % represents the deviation of the target value from the measured amplitude. In settled conditions, the display should come close to 0.0 % if the calibration is correct. If the error % is above 3 %, the device sounds an audible alert to indicate that the error limit is exceeded. In addition, the vibration calibrator switches off automatically in case of overload. Sensors can be connected via magnet or thread. A thread adapter is included in the delivery contents to accommodate M3, M5, M8, 1/4"-28 and UNF 10-32 threads.
The vibration calibrator has many versatile features for field and laboratory use, including a built-in rechargeable battery, battery-saving automatic power-off function, battery charge indicator and AC power adapter for charging. In addition, the supplied carrying case allows for convenient handling and safe transport.
HIGHLIGHTS
- Load-independent vibration magnitude
- Display shows frequency, amplitude, error percentage (%) and date of last calibration
- Sounds an alert if error limit exceeded
- Designed for use in the laboratory as well as the field
SPECIFICATION
Vibration magnitude (RMS values ± accuracy) | |||||||||
Vibration acceleration | 10 m/s² ± 3 % (0 ... 40 °C / 32 ... 100 °F) | ||||||||
Vibration velocity | 10 mm/s ± 3 % (0 ... 40 °C / 32 ... 100 °F) | ||||||||
Vibration displacement | 10 µm ± 3 % (0 ... 40 °C / 32 ... 100 °F) | ||||||||
Vibration frequency | 159.15 Hz ± 0.05 % (-10 ... 55 °C / 14 ... 130 °F) | ||||||||
Radian frequency | 1000/s ± 0.05 % (-10 ... 55 °C / 14 ... 130 °F) | ||||||||
Settling time | < 10 seconds | ||||||||
Level indication | Error percentage (%) display, above ± 3 % audible alert | ||||||||
Max. weight of test object for stated accuracy | 600 g / 1.32 lb | ||||||||
Dynamic force of vibration exciter | 10 N | ||||||||
Max. tightening torque of vibration exciter | 2 Nm | ||||||||
Operating humidity range | < 90 % at 30 °C / 86 °F, non-condensing | ||||||||
Power supply | Built-in NiMH rechargeable battery, 7.2 V / 1.6 Ah | ||||||||
Max. operating hours with rechargeable battery | Approx. 5 hours at a weight of 100 g / 0.22 lb | ||||||||
Charging time | Approx. 4 hours | ||||||||
Charging status | Bar indicator | ||||||||
Automatic power off | 10 minutes / 1 ... 30 minutes adjustable | ||||||||
Stray magnetic field at the vibration head | < 0.2 mT | ||||||||
Dimensions (L x W x H) | Approx. 100 x 100 x 120 mm / 3.9 x 3.9 x 4.7" | ||||||||
Weight | Approx. 2200 g / 4.9 lbs |
See user manual for complete technical specifications. |
DELIVERY SCOPE
1 x thread adapter (M3, M5, M8, 1/4"-28, UNF 10-32)
1 x battery-charging power adapter (100 ... 240 V AC; 50 / 60 Hz)
1 x factory calibration certificate
1 x user manual
1 x carrying case