Description
The TOMLIN INSTRUMENTS TMI-FM-108 is a high-accuracy portable transit-time ultrasonic flow meter for pipes from DN25 to DN5000 (1 to 200 inch). Rail-mounted clamp-on transducers keep the sensor spacing exact for repeatable readings.
Accuracy and linearity are ±0.5% from 0.5 to 5 m/s, with 0.15% repeatability, for check metering, audits and commissioning.
Features
- Portable transit-time ultrasonic meter
- Rail-mounted clamp-on transducers
- Pipes DN25 – 5000 (1 – 200 inch)
- Accuracy and linearity ±0.5%
- 3.5-inch IPS display
- Data logging
Specifications
| Model | TMI-FM-108 |
| Measuring principle | Transit-time ultrasonic, clamp-on |
| Pipe size | DN25 – DN5000 (1 – 200 inch) |
| Accuracy | ±0.5% (0.5 – 5.0 m/s) |
| Linearity | ±0.5% (0.5 – 5.0 m/s) |
| Repeatability | 0.15% |
| Flow velocity range | 0.1 – 12.0 m/s |
| Response time | 500 ms |
| Display | 3.5-inch 480 × 320 IPS colour |
| Data logging | Yes |
| Data storage | 10 years, 64 months, 64 days |
| Analogue output | 4 – 20 mA (max. load 600 Ω) |
| Communication | RS485, Modbus RTU |
| Pulse output | OCT |
| Power supply / battery | to be confirmed |
| Battery life | 5 – 6 years (rated; whether operating or standby life is to be confirmed) |
| Protection | Transmitter IP54; transducers IP68 |
| Transducer temperature | -40 – 80 °C |
| Transducer cable | 5 m standard (extendable to 20 m) |
Industries Used In
- Check metering and verification
- Flow audits on large mains
- Pump performance testing
- Commissioning of water systems
- Process and cooling water