Description
The TOMLIN INSTRUMENTS TMI-FM-109 is a compact clamp-on ultrasonic flow meter for small pipes from DN6 to DN100. The transmitter and transducer sit in one aluminium housing that clamps around the pipe with a magnetic bracket, so it fits in about a minute without cutting the pipe.
A colour IPS display shows flow rate, velocity and totals, and 4 – 20 mA and RS485 Modbus outputs connect to PLC and building systems. Accuracy is ±2.0% from 0.5 to 5 m/s on clean liquids in metal or plastic pipe.
Features
- Clamp-on, non-invasive: no pipe cutting or shutdown
- Magnetic latching bracket for one-person fitting
- Transmitter and transducer in one housing
- Gel-free coupling pads
- 4 – 20 mA and RS485 Modbus outputs
- Menu-driven 4-key setup
Specifications
| Model | TMI-FM-109 |
| Type | Compact clamp-on: transmitter and transducer in one housing |
| Pipe sizes | DN6 – DN100 (pipe OD 9 – 115 mm) |
| Liquid temperature | 0 – 50 °C |
| Accuracy | ±2.0% (0.5 – 5.0 m/s) |
| Flow velocity range | 0.1 – 5.0 m/s |
| Linearity | ±2.0% (0.5 – 5.0 m/s) |
| Repeatability | 0.8% |
| Response time | 500 ms |
| Display | 1.54-inch 240 × 240 IPS colour, four-way rotation |
| Units | m³/h, L/min, L/s, L/h, mL/min, GPM |
| Totaliser | 7-digit; net, single, day / month / year totals |
| Data storage | 10 years, 64 months, 64 days |
| Keypad | 4 touch keys |
| Power supply | 24 V DC, 3 W |
| Analogue output | 4 – 20 mA (max. load 600 Ω) |
| Communication | RS485, Modbus RTU |
| Signal output | Flow alarm, totaliser pulse (optional) |
| Protection | IP54 |
| Housing | Aluminium alloy |
| Ambient humidity | 0 – 95% RH, non-condensing |
Ordering Information
| TMI-FM-109-01 | DN6 (1/8"), pipe OD 9 – 15 mm |
| TMI-FM-109-02 | DN8 (1/4"), pipe OD 9 – 15 mm |
| TMI-FM-109-03 | DN10 (3/8"), pipe OD 15 – 21 mm |
| TMI-FM-109-04 | DN15 (1/2"), pipe OD 15 – 21 mm |
| TMI-FM-109-05 | DN20 (3/4"), pipe OD 21 – 28 mm |
| TMI-FM-109-06 | DN25 (1"), pipe OD 28 – 36 mm |
| TMI-FM-109-07 | DN32 (1 1/4"), pipe OD 36 – 44 mm |
| TMI-FM-109-08 | DN40 (1 1/2"), pipe OD 44 – 52 mm |
| TMI-FM-109-09 | DN50 (2"), pipe OD 52 – 66 mm |
| TMI-FM-109-10 | DN65 (2 1/2"), pipe OD 66 – 80 mm |
| TMI-FM-109-11 | DN80 (3"), pipe OD 80 – 95 mm |
| TMI-FM-109-12 | DN100 (4"), pipe OD 100 – 115 mm |
Industries Used In
- Water, process and cooling water lines
- Clean chemicals, acids and alkalis
- Plastic and metal pipe (PVC, PP-R, HDPE, steel, copper)
- Building services and HVAC
- Retrofit metering without cutting the pipe