Noise measuring device for fixed installation with relay output and control output
The condition monitoring sound level meter consists of a measuring microphone with handle and a wall holder with 1.5 m cable, a noise transmitter and a digital display. The main application of the noise meter is the continuous noise measurement and noise monitoring of factory halls, production rooms and large events (discotheques, folk festivals, kart tracks, etc.). The digital display has a relay output for the control of alarm horns or blinkers for visual or acoustic warning and a control output.
The limit value for this alarm output can be programmed on the display. The analogue output signal can also be used for controlling window flaps ... or the signals of the transmitter can be stored, evaluated and documented on a logger. The noise warning device is very easy to connect and can be operated via 240 V mains power. The cabling is the only installation task that the customer has to carry out himself. Thus, this noise warning device can be used quickly for monitoring purposes.
Specification
Sound sensor | |
Measuring range | 30 ... 130 dB in 3 ranges: I. 30 ... 80 dB(A) II. 50 ... 100 dB(A) III. 80 ... 130 dB(A) |
Resolution | 0.1 dB |
Accuracy | ±1.5 dB |
Frequency weighting | A |
Power supply | via transmitter |
Cable length (sensor to transmitter) | 1.5 m / 4.92' |
Wall mount | for screwing (included) |
Operating conditions | 0 ... +50 °C / +32 ... +122 °F, max. 85 % RH |
Microphone | 1/2" precision electret |
Standard | IEC 61 672 (Class II) |
Transmitter | |
Housing | ABS plastic |
Power supply | 90 … 260 ACV (max. 1.5 A) |
Calibration | adjustable by means of a screw |
Output | 4 ... 20 mA |
Protection class | IP 54 ingress protection rating |
Operating conditions | 0 ... +50 °C / +32 ... +122 °F, max. 85 % RH |
Digital indicator | |
Input signal from transmitter | 4 – 20 mA |
Indicator | -1999 ... 9999 (1 ... 3 decimals, according to the measurement value / chosen by the user) |
Programming data (decimals, limit value ..) | saved in the unit |
Regulation output | 3 points (Com, NO, NC) |
Alarm output | Relay output: 5 A / 240 VAC |
Power | 90 ... 260 ACV / 50 ... 60 Hz |
Dimensions / Switchboard according to DIN | 92 (+0.8) mm x 45 (+0.5) mm |
Weight | ca. 250 g / 0.56 lb |
Operating conditions | 0 ... +50 °C / +32 ... +122 °F, max. 85 % RH |
Protection class | IP 65 ingress protection rating |
Sound Calibrator PCE-SC 43 | |
Sound pressure level | 94 dB, 104 dB, 114 dB |
Accuracy | ±0.4 dB |
Class | 2 |
Frequency | 1000 Hz for A, B, C and D frequency weighting |
Frequency accuracy | ±1.7 % |
Stabilization time | 10 seconds |
Total harmonic distortion | <3 % |
Environmental conditions | |
Temperature influences | <0.4 dB at 0 ... 40 °C / 32 ... 104 °F |
Humidity influences | 25 ... 90 % RH |
Atmospheric pressure | <0.1 dB at 65 ... 108 kPa |
Stability at <60 seconds | ±0.15 dB |
Stability after one year | ±0.35 dB |
(normal use) | |
Operating conditions | 0 ... 40 °C / 32 ... 104 °F, <25 ... 90 % RH, |
non-condensing | |
Storage conditions | -20 ... 50 °C / -4 ... 122 °F, <90 % RH, |
non-condensing | |
Automatic shutdown | after 20 minutes |
Power supply | 2 x 1.5 V AA batteries |
Battery life | ca. 50 hours |
Microphone size | 1/2 inch |
Dimensions | 60 x 130 x 37.5 mm / 2.3 x 5.1 x 1.4" |
Weight | 400 g / 14.1 oz |
Delivery Scope
1 x Condition Monitoring Sound Level Meter PCE-SLT
1 x Noise sensor on 1.5 m / 4.92' cable
1 x Noise transmitter
1 x Digital indicator
1 x Sensor mount
1 x Instruction manual
1 x Sound Calibrator PCE-SC 43
1 x 9V battery
1 x Instruction manual
1 x Noise sensor on 1.5 m / 4.92' cable
1 x Noise transmitter
1 x Digital indicator
1 x Sensor mount
1 x Instruction manual
1 x Sound Calibrator PCE-SC 43
1 x 9V battery
1 x Instruction manual